Certificates and test reports
A qualified calibration of measuring devices is, especially in connection with companies' in house quality management systems, obtaining more and more significance. The traceability back to national measuring standards is hence very important if f. ex. measuring devices are used as reference.
BD|SENSORS therefore established in 1996 a nationally accredited metrological calibration laboratory belonging to ILAC (International Laboratory Accredition Cooperation). Consequently BD|SENSORS is able to perform calibration services according to internationally approved standards at the highest stage.
If requested DKD - now DAKKS calibration certificates can be delivered too.
Measurement reports
5-value-report
What is recorded in writing?
- measurement points: 5
- declaration of the measuring inaccuracy: no
- reference: IEC61298-2
- pressure cycles: 1x increasing / 1x decreasing
- traceability to nat. nomal: yes
- implementation: production plant
11-value-report
What is recorded in writing?
- measurement points: 11
- declaration of the measuring inaccuracy: no
- reference: IEC61298-2
- pressure cycles: 1x increasing / 1x decreasing
- traceability to nat. nomal: yes
- implementation: production plant
IEC-measurement report
What is recorded in writing?
- measurement points: 11
- declaration of the measuring inaccuracy: no
- reference: IEC61298-2
- pressure cycles: 3x increasing / 3x decreasing
- traceability to nat. nomal: yes
- implementation: production plant
Certificates of tests
Certificate of Compliance 2.1 (EN 10204)
What is certified?
BD|SENSORS confirms, that the delivered products comply with the order agreements. Test results are not documented.
Test report 2.2 (EN 10204)
What is certified?
BD|SENSORS confirms, that the delivered products comply with the order agreements. Specific event features (ranges, accuracy, etc.) are confirmed.
Inspection certificate 3.1 (EN 10204)
What is certified?
BD|SENSORS certifies, that during the production of the devices listed, the required material tests were conducted. The tests will be certified with expression of test results.
Calibration certificate
What is recorded?
- measuring points:
11 per test series (Process A, B - DKD guidline 6-1 03/2014)
6 per measuring series (Process C - DKD guidline 6-1 03/2014) - measuring series
2x increasing / 2x decreasing (Process A - DKD guidline 6-1 03/2014)
2x increasing / 1x decreasing (Process B - DKD guidline 6-1 03/2014)
1x increasing / 1x decreasing (Process C - DKD guidline 6-1 03/2014
- Declaration of measurement uncertainty: yes
- Reference: DKD guidline 6-1 03/2014
- Traceability to nat. standard: yes
- Realization: accredited calibration laboratory